• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE ASSOCIATES, INC. GORE VIATORR TIPS ENDOPROSTHESIS; SHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE ASSOCIATES, INC. GORE VIATORR TIPS ENDOPROSTHESIS; SHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS Back to Search Results
Model Number PTB8106275
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Event Description
During the deployment of a covered stent, the uncovered part of the stent deployed into the patient properly but then to finish deploying the covered part, the md needed to unscrew a plastic wheel which releases the covered part of the stent.In this case, the plastic wheel locked up and could not be opened, leaving the stent stuck in the patient half-deployed.The md was able to break the plastic wheel part off and grasp the string and deploy the stent properly.There was no harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIATORR TIPS ENDOPROSTHESIS
Type of Device
SHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Manufacturer (Section D)
W. L. GORE ASSOCIATES, INC.
1505 north fourth st
flagstaff AZ 86004
MDR Report Key12744164
MDR Text Key279828717
Report Number12744164
Device Sequence Number1
Product Code MIR
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPTB8106275
Device Catalogue NumberPTB8106275
Device Lot Number23175097
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/01/2021
Event Location Hospital
Date Report to Manufacturer11/03/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-