Device evaluated by mfr.: the device returned contained an angiojet solent omni.The assembly, supply line, shaft, tip, and spike line were all visually examined for potential damage.The catheter had multiple kinks along the catheter shaft.A functional test was performed.The pump was placed into the ultra drive console and the device primed with no complication.The device was run for a total time of 90 seconds in thrombectomy mode.The device stayed within the expected range of the product, with a pressure of 10.646 kpsi.There is no confirmation for the complaints allegation of shaft perforation or loss of aspiration.
|