Model Number N/A |
Device Problems
Failure to Cut (2587); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that before surgery the handpiece stuck and there was an irregular cut.No adverse event was reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor speed was unstable and motor was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
No additional event information.
|
|