Model Number CI-1500-01 |
Device Problems
Loose or Intermittent Connection (1371); Mechanical Problem (1384); Patient Device Interaction Problem (4001); Intermittent Communication Failure (4038)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2021 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however the issue has not resolved.An x-ray confirmed that the device is in the correct position.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Additional information: disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode array was severed.Slices in the silicone were also observed on the top cover.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed broken antenna wires.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device had broken antenna coil wires.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|