• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bradycardia (1751); Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226)
Event Date 10/14/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the patient became bradycardic and their blood pressure dropped.An ultrasound was performed and a cardiac tamponade was observed.The case was aborted.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Correction: d9 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the data files and 4fc12 sheath with lot number 0010499703 were returned and analyzed.The returned patient file showed at least two applications were performed with balloon catheter.No system notices were triggered during any of these applications.Visual inspection of the sheath before functional testing and dissection was performed on the shaft, handle, and dilator.No anomaly was identified during the external visual inspection.The handle, shaft and sideport were intact with no apparent issue.Visual inspection of the shaft area was performed.No anomalies were identified.The shaft is intact with no apparent issue.No kink or any damage was observed on the surface.The functional testing was performed and no anomaly was discovered.The performance test with sentinel blackbelt leak tester was performed.The pressure test with 30 psig showed the pressure decay in the device was 0.021 psig.The flushing test with 6 psig showed the pressure decay in the device was 0.001 psig.The aspiration test with negative pressure of 4.1 psig showed the pressure decay in the device was 0.023 psig.All the performance tests were in the acceptable range.The shaft, side tube, and valve were all leak-tight with no apparent issue.In conclusion, the reported tamponade, hypotension, and bradycardia issues are clinical events.They cannot be assessed through clinical data review.The sheath passed the returned product inspection as per specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12747197
MDR Text Key279949671
Report Number3002648230-2021-00503
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/23/2022
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010499703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2021
Initial Date FDA Received11/03/2021
Supplement Dates Manufacturer Received11/16/2021
11/16/2021
Supplement Dates FDA Received12/10/2021
12/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
990063-015 MAPPING CATHETER; 990063-015 MAPPING CATHETER
Patient Outcome(s) Life Threatening;
-
-