Catalog Number SGC0702
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Air Embolism (1697)
Event Date 10/19/2021
Event Type  Injury  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report the air embolism requiring intervention.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3.After removing the dilator from the inserted steerable guide catheter (sgc) an air bubble was noted in the left atrium (la).The sgc was steered directly above the air bubble and aspiration performed.The air bubble was removed and the procedure continued.One clip was implanted, reducing mr to 1.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have resulted in this event.Based on the information reviewed, a conclusive cause for air embolism could not be determined in this complaint.The reported patient effect of emboli (air embolism) as listed in the mitraclip g4 system instructions for use (ifu) is a known possible complication associated with mitraclip procedures.The reported unexpected medical intervention appears to be due to case specific circumstance as aspiration was performed to resolve the air embolism.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12749402
MDR Text Key280073747
Report Number2024168-2021-09973
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/02/2022
Device Catalogue NumberSGC0702
Device Lot Number10430R157
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2021
Initial Date FDA Received11/03/2021
Supplement Dates Manufacturer Received11/05/2021
Supplement Dates FDA Received11/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-