Mayfield skull clamp (a2000) was returned for evaluation: device history record (dhr): the dhr was reviewed and show no abnormalities related to the reported failure.Failure analysis: unit received with the hex bushing is worn and the lock still moves after unit is lock down.The hex bushing was replaced due to wear.The large starbursts was machined to add helicoils.General maintenance, cleaning and replacement of worn components performed.Root cause analysis: the reported complaint was confirmed from the evaluation.Evaluation showed that the hex bushing is worn and the lock still moves after unit is lock down.The unit needed repair, replacement of worn and damaged parts along with general maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|