MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH ROD, 480 MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH ROD, 480 MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 179762480
Device Problem Migration (4003)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/01/2021
Event Type  Injury  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Health effect: clinical code (b)(4) used to capture: injury.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that, post-operatively the rods slide out of the distal (caudal) screws, despite of a double control with the torque handle.The surgeon does not suspect a defect with the torque handle.This report is for one (1) expedium spine system rod 5.5 x 480mm.This is report 1 of 2 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was further reported that the surgeon decided to keep the rod as it is.Patient outcome is reported as okay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD, 480 MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12759361
MDR Text Key280753643
Report Number1526439-2021-02329
Device Sequence Number1
Product Code OSH
UDI-Device Identifier10705034077437
UDI-Public(01)10705034077437
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K111136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179762480
Device Catalogue Number179762480
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2021
Initial Date FDA Received11/07/2021
Supplement Dates Manufacturer Received11/16/2021
Supplement Dates FDA Received12/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ROD, 480 MM; UNKNOWN TORQUE DEVICES
-
-