It was reported that the procedure was to treat the carotid artery with 90% stenosis.The emboshield nav6 embolic protection system (eps) was released in the intended location without issue; however, the delivery catheter could not be withdrawn.Therefore, the emboshield nav6 was removed with the delivery catheter and a new emboshield nav6 was used to complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
|
The device was returned for analysis.The reported difficult to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the 90% stenosed anatomy and/or manipulation of the device resulted in the noted device damages (bent delivery catheter shaft, bent barewire-core, bunched/wavy barewire-coils) thus resulting in the reported difficult to remove.Manipulation of the device likely resulted in the noted wrinkled dc pod.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|