One 4 lesion nt2000¿ pain management rf generator was received to service.Inspection of the rf control board showed burnt components near tr28.Additionally, tp14 was damaged, resulting in the inability of capacitance readings to be taken from that location.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Based on the information provided to abbott and the investigation performed, the root cause of the field reported issue was isolated to abnormal functionality of the radio frequency control board.
|