Model Number RFG-NT-2000
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One 4 lesion nt2000¿ pain management rf generator was received to service.Inspection of the rf control board showed burnt components near tr28.Additionally, tp14 was damaged, resulting in the inability of capacitance readings to be taken from that location.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Based on the information provided to abbott and the investigation performed, the root cause of the field reported issue was isolated to abnormal functionality of the radio frequency control board.
 
Event Description
During service of the system, the rf control board had burnt components near tr28.The rf board was replaced to resolve the issue.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR
Type of Device
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12759724
MDR Text Key280439643
Report Number2184149-2021-00350
Device Sequence Number1
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111576
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Physician
Type of Report Initial
Report Date 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number6558242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/25/2021
Initial Date FDA Received11/07/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/31/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-