SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMING DEVICE; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported the device had "temperature issues".Issue was found during preventive maintenance; no patient involved.
|
|
Manufacturer Narrative
|
Additional information: d4 udi is unknown.No product information has been provided to date.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.Pole clamp color was faded.Started with a visual inspection then filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The reported issue was confirmed the overtemp setting was out of calibration.The root cause of the reported issue could not be determined.The technician reset the overtemp during calibration.
|
|
Event Description
|
Additional info received via email 01-nov-2021: date of event is unknown, was tested before patient use, adverse effects n/a.
|
|