The patient was undergoing a coil embolization procedure to treat a pulmonary arteriovenous malformation (pavm) using a lantern delivery microcatheter (lantern), ruby coils and a non-penumbra guidewire.During the procedure, the physician experienced difficulty advancing the lantern over the guidewire and subsequently, the guidewire would not advance further.Therefore, the physician removed the lantern and guidewire.Upon removal, it was noticed that the guidewire perforated the lantern at the mid-shaft.The procedure was completed using a new lantern, twelve ruby coils and a new non-penumbra guidewire.There was no report of an adverse effect to the patient.
|
Evaluation of the returned lantern revealed that the catheter was fractured.This is likely the reported perforated location.If the lantern is mishandled at an extreme angle while advancing a guidewire through its lumen, damage such as this may occur.Further evaluation of the revealed a kink.This damage is likely incidental to the complaint.Evaluation of the returned non-penumbra guidewire revealed a functional device.The guidewire was advanced through a demonstration microcatheter without an issue.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|