• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PRSTHS, KNEE, PATFEMIB, SEMI-CNSTRND, CMNTD, POLY/MET/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PRSTHS, KNEE, PATFEMIB, SEMI-CNSTRND, CMNTD, POLY/MET/POLY Back to Search Results
Model Number 74012812
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Event Description
It was reported that, during spd cleaning, the attachment arm of the fem implant impact was noticed cracked.As this was noticed in a non-surgical environment, there was not any patient involved.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned impactor confirms the cam arm has a crack in it, rendering the device inoperable.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRN II BCS LCK FEM IMPLANT IMPACT
Type of Device
PRSTHS, KNEE, PATFEMIB, SEMI-CNSTRND, CMNTD, POLY/MET/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12763406
MDR Text Key282581592
Report Number1020279-2021-07878
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556003541
UDI-Public00885556003541
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111711
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74012812
Device Catalogue Number74012812
Device Lot Number18CST0005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/15/2021
Initial Date FDA Received11/07/2021
Supplement Dates Manufacturer Received11/22/2021
Supplement Dates FDA Received11/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-