• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU Back to Search Results
Model Number DM0008FAA
Device Problem Device Remains Activated (1525)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2021
Event Type  malfunction  
Event Description
It was reported that an easydrill training was organized and conduct a surgical technique using a 3d skull model.They drilled several holes with different perforators and out of the 18 holes made using two dm0008faa (l/n: 482/20), only one perforator created 4 plunges.It also reported that several holes were made using dm0011faa (l/n: 462/20) and some holes plunges through the 3d model.It was also reported that there was no patient involvement.
 
Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.This dm0008faa easydrill cranial perforator with lot number 482/20 and dm0011faa easydrill cranial perforator with lot number 462/20 were manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR  09980-39
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR   09980-39
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--39
BR   09980-39
140575722
MDR Report Key12763461
MDR Text Key286643670
Report Number1625507-2021-00350
Device Sequence Number1
Product Code HBF
UDI-Device Identifier07898959543166
UDI-Public07898959543166
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date04/22/2024
Device Model NumberDM0008FAA
Device Catalogue NumberDM0008FAA
Device Lot Number482/20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DM0011FAA (L/N: 462/20)
-
-