• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY USPO2 PULSE OXIMETRY CABLE KIT; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY USPO2 PULSE OXIMETRY CABLE KIT; OXIMETER Back to Search Results
Model Number 3577
Device Problem Sensing Intermittently (1558)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2021
Event Type  malfunction  
Manufacturer Narrative
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
The customer reported shortages in the cord then it stopped working over the weekend.No patient impact or consequences were reported.
 
Event Description
The customer reported shortages in the cord then it stopped working over the weekend.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.Visual inspection found no external damage or defects.Device functionality testing found the unit established communication with interfacing test fixture; however, the module does not communicate properly with the connected sensor.Device was observed through x-ray imaging and no defects were observed.The device is sealed and cannot be opened for internal inspection without damaging the module.Exact component failure could not be determined due to device being fully potted.A service history record review reveals that this unit was in the field for over five (5) months with no previous reported issues related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
USPO2 PULSE OXIMETRY CABLE KIT
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key12763540
MDR Text Key283303130
Report Number3019388613-2021-00306
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997007481
UDI-Public00843997007481
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121914
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3577
Device Catalogue Number3577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2021
Was the Report Sent to FDA? No
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-