For a vsp reconstruction procedure, the surgeon identified the following issues: the vsp reconstruction surgical plan documented on the 3d systems case report provided to the surgeon did not align with the virtual simulated plan.The surgical procedure was documented as left fibula to right neck; instead of right fibula to right neck.This issue was identified by the surgeon prior to surgery.3d systems verified that the manufactured anatomical fibula cutting guide was designed to the correct right fibula right neck surgical plan; and the issue was isolated to a case report documentation error.3d systems corrected the case report and submitted a new revision to the surgeon prior to surgery.The surgical outcome was successful with no reported adverse effects to the patient.
|