Zoll has not received the product for investigation.A follow-up report will be submitted when the product is returned and investigation has been completed.According to available information, the death was not related to the autopulse device.The autopulse is used as an adjunct to manual cpr in cases of clinical death.The benefit of using the autopulse is that it in part substitutes mechanical compressions for the physical labor of manual chest compressions.If the autopulse did not start or unexpectedly stops compressions, rescuer should revert to manual cpr, which is the standard of care.In this case, the autopulse platform (sn unknown) was used to resuscitate patient who was into a cardiac arrest for approximately 10 minutes.The autopulse platform (sn unknown) displayed an error message after three compressions, and then the platform completely shut down.At this point, manual compressions were resumed, troubleshooting was unsuccessful.According to medwatch report, the patient was dead.It's unknown if return of spontaneous circulation (rosc) was once achieved or not during the resuscitation per available information.The autopulse is intended to be used as an adjunct to manual cpr on adult patients.In case of stoppage of autopulse the trained user reverts to manual cpr.The transition from autopulse to manual cpr by trained users is similar to the time necessary for rescuer rotation, and presents the same workflow as manual cpr.Hence, based on available information, the patients' outcome was not negatively impacted by the interruptions when compared to standard of care manual cpr.Out-of-hospital cardiac arrest (ohca) is one of the main causes of death in industrial nations.About 25% of patients survive this event and make it to the hospital, and even fewer patients survive after 24 hours (nichol, nejm, 2015).In the united states, survival to hospital discharge after non-traumatic emergency medical services-treated cardiac arrest with any first recorded rhythm was 10.6% for patients of any age.Of the bystander-witnessed out-of-hospital cardiac arrests in 2011, 31.4% of victims survived to hospital discharge (mozaffarian, circulation, 2016).Death is an expected outcome for ohca.
|
During patient use, the autopulse platform (sn unknown) performed three compressions and the platform completely shut down.The customer performed manual cpr.Also noted, the slide gate on the battery compartment is broken.The three autopulse li-ion batteries were fully charged prior to the call.A return of spontaneous circulation (rosc) was not achieved, and the patient expired.The customer did not provide information regarding the relationship between the death and the alleged malfunction.However, zoll's medical safety assessment (msa) evaluated the incident, and it was determined that the death was not related to the autopulse device.
|