OLYMPUS MEDICAL SYSTEMS CORP. OES CHOLEDOCHOFIBERSCOPE; CHOLEDOCHO FIBERSCOPE Back to Search Results
Model Number CHF-CB30L
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to sorc but has not been returned to omsc.Sorc checked the subject device for evaluation.It was confirmed that the insertion tube of the subject device was peeled off more than 1mm2 due to the deterioration.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.It could not be identified the cause of the reported phenomenon.However based on the report of sorc, omsc surmised there was the possibility this phenomenon was attributed to the stress when using, the external factors, or the user handling.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center (sorc), it was found that the insertion tube of the subject device was peeled off more than 1mm2.The subject device had been returned to sorc for repair of the leakage at the bending section.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CHOLEDOCHOFIBERSCOPE
Type of Device
CHOLEDOCHO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12764077
MDR Text Key282052603
Report Number8010047-2021-14201
Device Sequence Number1
Product Code FBN
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCHF-CB30L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/06/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-