• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX200H11C
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Visual Impairment (2138); Eye Infections (4466)
Event Date 09/22/2021
Event Type  Injury  
Event Description
The manufacturer became aware of an allegation from a user that air was blown into the user's right eye from the user's mask becoming dislodged during cpap therapy.The user reports having cataract surgery in her right eye two days prior to the alleged date of event.The user reports a subsequent loss of vision in the right eye as well as diagnosis of an eye infection treated with surgery and antibiotics in the right eye.It was reported that, at the time of the event, the user had no vision in the right eye but at the date of this report, vision has returned about forty percent.The manufacturer has requested the return of the machine, mask and headgear for investigation.The manufacturer's investigation is ongoing.Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice /recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an allegation of an issue related to sound abatement foam that became degraded and caused the patient to have right eye go blind because the nasal mask blow air in her right eye.Based on the available information, the manufacture concludes no further action is necessary.There was no medical intervention required by the patient.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.  section h6 updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key12764283
MDR Text Key280604132
Report Number2518422-2021-06366
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberDSX200H11C
Device Catalogue NumberDSX200H11C
Was Device Available for Evaluation? No
Date Manufacturer Received08/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-