BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 1X40-3; ORTHOPAEDIC CEMENT, MEDICATED Back to Search Results
Catalog Number 3003940001-3
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Arrest (1762)
Event Date 10/11/2021
Event Type  Injury  
Manufacturer Narrative
This is a combination product (b)(4).Report source, foreign, health professional - event occurred in (b)(6).The device manufacturing quality record indicate that the released product met all requirements to perform as intended.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that: a patient was operated on (b)(6) 2021 for a left intermediate prosthesis.He had several intraoperative cardiac arrests, and went in the intensive care unit.Poor prognosis , the patient is in intensive care.
 
Event Description
It was reported that the patient was operated on (b)(6) 2021 for a left intermediate prosthesis.He had several intraoperative cardiac arrests, and is in the intensive care unit, with poor prognosis.
 
Manufacturer Narrative
This is a combination product (b)(4).This supplemental report is to relay additional information.The product analysis can't be performed as the product was not returned.The device manufacturing quality record indicate that the released product met all requirements to perform as intended.No non conformity was found on the reference and batch number.1 other complaint on patient death following bone cement implantation syndrom was recorded on the batch since ever, and 2 other complaints on patient death following bone cement implantation syndrom was recorded from (b)(6) 2018 to (b)(6) 2021.Please note that all events occurred at the same hospital.The ifu of the product has been reviewed and it contains risk -reducing warning and different precautions regarding severe complications): "after preparation of the prosthesis bed or directly after the implantation of the cement and prosthesis, pressure rise in the medullary canal may cause a temporary fall in blood pressure.In addition to hypotension, pulmonary embolism and cardiac arrest, with their potentially fatal consequences, have been encountered in rare cases".According to available data, root cause of the event was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient was operated on (b)(6) 2021 for a left intermediate prosthesis.Had several intraoperative cardiac arrests.The patient is in intensive care with poor prognosis.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Product from the same batch number and retained by the hospital has been received for analysis.The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.A sample returned from the hospital of same batch has been tested in the laboratory under standardized conditions.No unusual behavior during mixing, handling or setting.The device manufacturing quality record indicate that the released product met all requirements to perform as intended.The review of the raw material certificate shows no non conformity or deviation.Medical records were reviewed and there was no information regarding patient health condition that was not previously known.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN BONE CEMENT R 1X40-3
Type of Device
ORTHOPAEDIC CEMENT, MEDICATED
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12766469
MDR Text Key280455813
Report Number3006946279-2021-00190
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number3003940001-3
Device Lot NumberZ20BAD1401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/13/2021
Initial Date FDA Received11/08/2021
Supplement Dates Manufacturer Received12/17/2021
05/31/2022
07/25/2022
Supplement Dates FDA Received12/17/2021
06/27/2022
08/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-