MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2021
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver did not pass the system check out.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that freedom driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found two new alarms, indicating secondary motor voltage too high and speaker 2 voltage too high when on, recorded in driver's data file.Visual inspection of external components found cosmetic damage to the display window and the display and fan covers.Visual inspection of internal components found top left anchor boss pulled out and top righa and bottom left anchor boss cracked.Scuffs were found within in the piston cylinder assembly as well as a melt mark on the ribbon cable that runs from the speaker to lcd near primary motor.Freedom driver passed all areas of functional testing for acceptance at incoming inspection.Additional testing included an extended observation run and separate testing of the secondary motor system to determine functionality.No alarms or malfunctions were observed.Primary and secondary motors were found to spin by hand freely.Complaint could not be replicated.Failure investigation for this complaint confirmed the reported issue.The customer complaint was not replicated during testing; root cause of the unresolvable red alarm was unable to be determined.Failure investigation identified no test failures or damage that could have contributed to the complaint.Cracks and scuffs found internally and externally are cosmetic only.No device malfunction found.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver did not pass the system check out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key12771397
MDR Text Key281417245
Report Number3003761017-2021-00212
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-