Model Number 595000-001 |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver did not pass the system check out.
|
|
Manufacturer Narrative
|
Device history record (dhr) review confirmed that freedom driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found two new alarms, indicating secondary motor voltage too high and speaker 2 voltage too high when on, recorded in driver's data file.Visual inspection of external components found cosmetic damage to the display window and the display and fan covers.Visual inspection of internal components found top left anchor boss pulled out and top righa and bottom left anchor boss cracked.Scuffs were found within in the piston cylinder assembly as well as a melt mark on the ribbon cable that runs from the speaker to lcd near primary motor.Freedom driver passed all areas of functional testing for acceptance at incoming inspection.Additional testing included an extended observation run and separate testing of the secondary motor system to determine functionality.No alarms or malfunctions were observed.Primary and secondary motors were found to spin by hand freely.Complaint could not be replicated.Failure investigation for this complaint confirmed the reported issue.The customer complaint was not replicated during testing; root cause of the unresolvable red alarm was unable to be determined.Failure investigation identified no test failures or damage that could have contributed to the complaint.Cracks and scuffs found internally and externally are cosmetic only.No device malfunction found.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
|
Event Description
|
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver did not pass the system check out.
|
|
Search Alerts/Recalls
|