There was an allegation of a display issue with a coaguchek xs meter.The reporter stated the meter's batteries have been replaced three times within the last three weeks.The time and date had to be set each time the batteries were replaced.The reporter attempted to run a test, and an error 5 appeared in the center of the meter's display.The reporter stated on the top of the 5, there were three dashes and then a dash on the left side and right side of the 5.The reporter confirmed the bell was flashing in the top right corner of the display.The reporter performed a display check and confirmed there were no missing segments.The reporter entered the meter's memory and had no issues with the last test result of 2.5 inr on (b)(6) 2021.The reporter confirmed the battery well is clean, dry, and not corroded.Also, the reporter confirmed the test strip guide was clean.
|