Model Number LN130B
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the vent valve leaked.There was a 5 minute delay while the pack was swapped out.Product was changed out.Procedure was completed successfully.There was a blood loss of 10-20cc.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 9, 2021.  all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
Additional information received that the line used was, the vent line that is connected to y¿d antegrade cardioplegia cannulae placed in the proximal ascending aorta.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer) ; h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The complaint sample was visually inspected upon receipt and a gap was noticed between the housings.It is unknown how or when a gap formed in the housing.All ops valves are subjected to a 100% leak test that undergoes 5 separate programs to test that each component in the valve is functioning properly.A retention sample was obtained and leak tested.It was found to function as intended.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE, O.P.S BULK, N-S
Type of Device
OVERPRESSURE SAFETY VALVE
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key12777917
MDR Text Key281682837
Report Number1124841-2021-00253
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLN130B
Device Catalogue NumberN/A
Device Lot NumberZG14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-