TERUMO CARDIOVASCULAR SYSTEMS CORPORATION VALVE, O.P.S BULK, N-S; OVERPRESSURE SAFETY VALVE
|
Back to Search Results |
|
Model Number LN130B |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
|
Event Description
|
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the vent valve leaked.There was a 5 minute delay while the pack was swapped out.Product was changed out.Procedure was completed successfully.There was a blood loss of 10-20cc.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 9, 2021. all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Event Description
|
Additional information received that the line used was, the vent line that is connected to y¿d antegrade cardioplegia cannulae placed in the proximal ascending aorta.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer) ; h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The complaint sample was visually inspected upon receipt and a gap was noticed between the housings.It is unknown how or when a gap formed in the housing.All ops valves are subjected to a 100% leak test that undergoes 5 separate programs to test that each component in the valve is functioning properly.A retention sample was obtained and leak tested.It was found to function as intended.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|