• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ3 7MM RM/LL; EARLY INTERVENTION : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ3 7MM RM/LL; EARLY INTERVENTION : KNEE TIBIAL INSERT Back to Search Results
Model Number 1024-54-307
Device Problems Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Patient Problems Fatigue (1849); Pain (1994); Loss of Range of Motion (2032); Insufficient Information (4580)
Event Date 10/19/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The poly had worn through.The poly is available to collect from the hospital however it is contaminated and will not be decontaminated by the local sterilization hub.A poly insert exchange was completed no further action for the patient.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was received for examination.Also, available x-rays were reviewed, and evidence of implant disassociation was found.After reviewing physical condition of the part there was observed a remarkable worn condition throughout the part consistent with localized wear due to the disassociation of the insert from the uni tibial tray.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ3 7MM RM/LL
Type of Device
EARLY INTERVENTION : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12779024
MDR Text Key280522175
Report Number1818910-2021-24635
Device Sequence Number1
Product Code HRY
UDI-Device Identifier10603295002833
UDI-Public10603295002833
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K070267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2017
Device Model Number1024-54-307
Device Catalogue Number102454307
Device Lot Number275553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2021
Initial Date FDA Received11/09/2021
Supplement Dates Manufacturer Received12/01/2021
12/13/2021
12/14/2021
01/21/2022
Supplement Dates FDA Received12/02/2021
12/14/2021
01/06/2022
01/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SIGMA HP UNI FEMORAL SZ3 RM/LL; SIGMA HP UNI INS SZ3 7MM RM/LL; SIGMA HP UNI TIB TRAY SZ3 RMLL
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-