SOCLEAN INC. SOCLEAN 2; SLEEP EQUIPMENT MAINTENANCE SYSTEM Back to Search Results
Model Number SC1200
Device Problem Decrease in Pressure (1490)
Patient Problems Fatigue (1849); Unspecified Respiratory Problem (4464)
Event Date 05/21/2021
Event Type  Injury  
Manufacturer Narrative
File a 30-day mdr.An assessment was conducted.The event is still considered reportable under fda's regulation.Soclean has reviewed and made improvements to our complaint handling procedure to ensure compliance with fda regulations.These improvements were recommended as a result of an internal audit conducted by an external independent regulatory consultant.We revised our decision trees and re-evaluated retrospective complaints to ensure reportable events were correctly identified.Per fdas guidance soclean is now reporting these retrospective mdrs.In addition, soclean has opened several capa's to remediate the audit findings.
 
Event Description
Customer reports being sick for 12 weeks with exhaustion and change in breathing.Requiring er visit with pulminary and cardiac work up in which were negative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCLEAN 2
Type of Device
SLEEP EQUIPMENT MAINTENANCE SYSTEM
Manufacturer (Section D)
SOCLEAN INC.
12 vose farm road
peterborough NH 03458
Manufacturer (Section G)
SOCLEAN INC.
12 vose farm road
peterborough NH 03458
Manufacturer Contact
anne nadeau
12 vose farm road
peterborough, NH 03458
6033712533
MDR Report Key12783454
MDR Text Key280574526
Report Number3009534409-2021-00073
Device Sequence Number1
Product Code LRJ
UDI-Device Identifier00187293000860
UDI-Public187293000860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberSC1200
Device Catalogue NumberSC1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2021
Initial Date Manufacturer Received 05/24/2021
Initial Date FDA Received11/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-