Model Number SC1200
Device Problem Insufficient Information (3190)
Patient Problem Cough (4457)
Event Date 08/16/2021
Event Type  Injury  
Manufacturer Narrative
File a 30-day mdr.An assessment was conducted.The event is still considered reportable under fda's regulation.Soclean has reviewed and made improvements to the complaint handling procedure to ensure compliance with fda regulations.These improvements were recommended because of an internal audit conducted by an external independent regulatory consultant.Soclean opened several capas to remediate the audit findings.Processes were revised to ensure reportable events were correctly identified and submitted to fda.Per fdas guidance, soclean is now reporting these events, which also includes retrospective items.Soclean believes this complaint is related to the philips recall of the dreamstation 1, not the soclean device.Soclean is processing this complaint in accordance with its complaint handling and quality system processes.
 
Event Description
Customer reports a cough with md intervention requiring antibiotics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCLEAN 2
Type of Device
SLEEP EQUIPMENT MAINTENANCE SYSTEM
Manufacturer (Section D)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer (Section G)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer Contact
anne nadeau
12 vose farm road
peterborough, NH 03458
6033712533
MDR Report Key12787254
MDR Text Key284857093
Report Number3009534409-2021-00371
Device Sequence Number1
Product Code LRJ
UDI-Device Identifier00187293000860
UDI-Public187293000860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberSC1200
Device Catalogue NumberSC1200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/16/2021
Initial Date FDA Received11/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-