Product analysis: the hawkone device was returned to the medtronic investigation lab for evaluation.The device was returned coiled inside four biohazard bags.The device was received with a bend on the strain relief, and the thumbswitch was fully advanced in the ¿off¿ positions.The cutter was positioned approximately 27mm inside the housing.A visual inspection found damage to the distal end of the cutter window.The cutter driver was activated and the thumbswitch was retracted to the ¿on¿ position and the cutter was able to fully return to the cutter window and rotate.No further testing could be carried out as the drive shaft broke.If information is provided in the future, a supplemental report will be issued.
|