Model Number N/A
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/07/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the insulation on the cable is damaged and the motor is not stable.The unit will be returned unrepaired as customer has not approved repair quote.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.It was noted the device has not been returned regularly for recommended annual pm.The event is confirmed.
 
Event Description
It was reported that the dermatome broke while being cleaned by the nurses.There was no patient involvement as the event took place outside of surgery.While cleaning the dermatome handpiece, the staff member pulled too hard on the electrical cord, where it attaches to the handpiece, and the wires came loose.Investigation found the motor speed was unstable.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12804254
MDR Text Key280702696
Report Number0001526350-2021-01269
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number62976464
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received11/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-