• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMING DEVICE; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMING DEVICE; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation- the device was returned for evaluation.The device was powered on and the display showed liquid crystal leakage.The cause of the component failure was not definitely established.
 
Event Description
It was reported the device display could not be read.Issue was found during preventive maintenance; no patient involved.
 
Manufacturer Narrative
Additional information b5 d5 h6.D4 udi is unknown.No product information has been provided to date.This mdr was generated under protocol b10010116, as a result of warning letter cms# 617147.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Wear and tear damaged enclosure and front cover.Cracked tank cover.Started with a visual inspection then filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The reported problem was confirmed.There was liquid crystal leakage in the liquid crystal display.The root cause of reported issue is dropping of the device by customer.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped., corrected data: d2 g1 g5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMING DEVICE
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
main receiving dock
minneapolis, MN 55442
7633833310
MDR Report Key12808267
MDR Text Key280763054
Report Number3012307300-2021-10635
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-