Additional information b5 d5 h6.D4 udi is unknown.No product information has been provided to date.This mdr was generated under protocol b10010116, as a result of warning letter cms# 617147.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Wear and tear damaged enclosure and front cover.Cracked tank cover.Started with a visual inspection then filled tank with water, attached temperature check, plugged in line cord, and turned on the power switch.The reported problem was confirmed.There was liquid crystal leakage in the liquid crystal display.The root cause of reported issue is dropping of the device by customer.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped., corrected data: d2 g1 g5.
|