Model Number H1-M |
Device Problems
Detachment of Device or Device Component (2907); Difficult to Advance (2920)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician was attempting to use the hawkone atherectomy device during procedure to treat severely calcified lesion in the mid superficial femoral artery(sfa).The vessel had no tortuosity and exhibited cto (chronic total occlusion-100%).Ifu was followed.It was reported that the tip detached at the nosecone.Resistance was felt during advancement.The procedure was completed without directional atherectomy.No patient injury.
|
|
Manufacturer Narrative
|
Additional information: the device could not make the acute turn through the tour guide into the left iliac.Unable to advance forward, decision was made to withdraw the hawkone catheter and settle for balloon angioplasty of the left sfa.It was reported that the tip detached at the nosecone abruptly upon pulling the hawkone catheter back.The shaft with cutter wheel intact was removed.The tip detached at the hinge point.No guidewire prolapse was noted.Fluoroscopy confirmed the nosecone fragment lodged at the curve of the tourguide, with approximately 7mm of the distal tip protruding from the tip of the sheath with the wire still in place.The detachment occurred within a 7french tour guide stealth.The detached component was removed from patient.The detached tip was snared and removed from the body.A microsnare was advanced and able to pulled a coiled metallic portion from the back of the nosecone, but was unable to loop any higher on the hawkone fragment.Under fluoroscopy visualization, the tour guide sheath was carefully withdrawn from the artery and the nosecone was removed from the body intact without any arterial embolization of fragments.Updated section 2 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: the hawkone device was returned to medtronic investigation lab for evaluation.The device was returned coiled inside four biohazard bags along with detached tip and pouch.A visual inspection showed that the tip detached distal to the anchor pockets, not at the hinge as reported.There is a bend on the detached tip beside the flush window and there is biologics inside the tip.A visual inspection of the detached tip found that the proximal section of the tip is missing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|