DEPUY CMW - 9610921 SMARTSET GMV 40G US EO; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 5450-50-501
Device Problem Loss of or Failure to Bond (1068)
Patient Problem Insufficient Information (4580)
Event Date 10/25/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Dmf# - 13704; trade name gentamicin sulphate; active ingredient(s) gentamicin sulphate; dosage form - powder; strength 1.0g active in our cements.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient was revised due to tibial tray loosening at the cement to implant interface.Two depuy cements were used.Doi: (b)(6) 2010; dor: (b)(6) 2021; left knee.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: dhr referenced from (b)(4).Device history reviewed for smartset gmv 40g us eo (b)(4), lot:2980913: no non conformances on this batch.Final micro and sterility tests passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GMV 40G US EO
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12810975
MDR Text Key280776092
Report Number1818910-2021-25086
Device Sequence Number1
Product Code MBB
UDI-Device Identifier10603295174295
UDI-Public10603295174295
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K081163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5450-50-501
Device Catalogue Number545050501
Device Lot Number2980913
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MBT CEM KEEL TIB TRAY SZ4; PFC SIGMARP STB TB IN 4 12.5; PFC*SIGMA/OV/DOME PAT 3PEG,35; SIGMA PS CEM FEM SZ4N L; SMARTSET GMV 40G US EO; SMARTSET GMV 40G US EO
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-