• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC GYRUS, PK-SP GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC GYRUS, PK-SP GENERATOR Back to Search Results
Model Number 744000
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device inspection found the core of the transformer on pkrf board was observed to be broken.The two mounting feet was found missing.Scratched were noted on the top case.Current software is (b)(4).Burn in test was performed and unit passed the test.Review of fault log shows 200 ref 92 one time: electrode id failure.Cpu checks for overactive id count flag not set.Cpu checks id count < 1500.Check accessories.This error can be caused by faulty test boxes or operator error and error 400 ref 26, six times: generated if a turis electrode is attached and activated without a saline solution being present, or there is an electrode connection fault.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
During an asset return inspection for device inspection and maintenance the core of the transformer on pkrf board was found broken.There is no patient involvement on this reported event.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the core transformer of the pkrf board was broken, though there was no report of device malfunction.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYRUS, PK-SP GENERATOR
Type of Device
GYRUS, PK-SP GENERATOR
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
118 turnpike road
southborough, MA 01772
9013785969
MDR Report Key12814064
MDR Text Key281933696
Report Number3003790304-2021-00163
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009271
UDI-Public00821925009271
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number744000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-