Model Number 744000 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Device inspection found the core of the transformer on pkrf board was observed to be broken.The two mounting feet was found missing.Scratched were noted on the top case.Current software is (b)(4).Burn in test was performed and unit passed the test.Review of fault log shows 200 ref 92 one time: electrode id failure.Cpu checks for overactive id count flag not set.Cpu checks id count < 1500.Check accessories.This error can be caused by faulty test boxes or operator error and error 400 ref 26, six times: generated if a turis electrode is attached and activated without a saline solution being present, or there is an electrode connection fault.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
|
Event Description
|
During an asset return inspection for device inspection and maintenance the core of the transformer on pkrf board was found broken.There is no patient involvement on this reported event.No user injury reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the core transformer of the pkrf board was broken, though there was no report of device malfunction.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|
|
Search Alerts/Recalls
|