Model Number 71174934
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2021
Event Type  malfunction  
Event Description
It was reported that, during an internal fixation surgery, the 2.0/2.7mm double-ended drill guide bent and became lodged with the sleeve.Also, the sleeve became bent and unable to drill through.Surgery was resumed, with greater than 30 minutes delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Additional information: d3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0/2.7MM DOUBLE-ENDED DRILL GUIDE
Type of Device
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12814979
MDR Text Key280825964
Report Number1020279-2021-08150
Device Sequence Number1
Product Code FZX
UDI-Device Identifier00885556387993
UDI-Public00885556387993
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71174934
Device Catalogue Number71174934
Device Lot Number17JTD0002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/21/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-