The subject device was received and evaluated.Evaluation found the reported issue of ¿would not turn on¿ was confirmed.The device was found with faulty psu (power supply unit).Using a reference psu, the unit started up and display showed serial # 0000000.Further inspection found unit missing 2 mounting feet, fan does not work and d socket was missing.In addition, during testing, the unit failed hand switch test rh #9 due to faulty auxiliary board and the output power at pbdes 200w, 300r over limit> 50mv due to faulty plasma blend board.The unit is running an old version software v2.06 and needs to be upgraded to v2.07.The housing has multiple minor scratches can use as is.Review of fault log shows 400 ref 20, two times: front panel: standby button stuck.The usual cause is that the relevant button is held down during power on self test or there is a faulty membrance keypad.600 ref 14, two times: footswitch mode pedal stuck.The usual cause is that the relevant foot pedal is held down by the user or there is a faulty foot pedal.100 ref 10, one time: software execution failure (watchdog reset).The generator software routines were interrupted unexpectedly by a watchdog reset) this can be cause by power¿ brown outs¿ or internal faults.Based on evaluation findings the reported failure was found to be attributed to faulty psu unit.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|