• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 II ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 II ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G2
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
The initial reporter received questionable elecsys ft4 ii results for one patient tested on a cobas 6000 e601 module with serial number (b)(4).The initial result was reported to the clinician, who asked that the sample be rerun on an abbott instrument.The patient's sample had an initial result of 2.24 ng/dl.On (b)(6) 2021, the repeat result was 0.88 ng/dl.
 
Manufacturer Narrative
The alarm trace, pre-analytical and calibration data did not indicate an issue.The investigation did not identify a general reagent problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 II ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12821522
MDR Text Key288023581
Report Number1823260-2021-03386
Device Sequence Number1
Product Code CEC
UDI-Device Identifier04015630934232
UDI-Public04015630934232
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G2
Device Catalogue Number06437281160
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2021
Initial Date FDA Received11/16/2021
Supplement Dates Manufacturer Received12/01/2021
Supplement Dates FDA Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-