Model Number CI-1601-05
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Tinnitus (2103); Vertigo (2134)
Event Type  Injury  
Event Description
The recipient is reportedly experiencing tinnitus and vertigo with device use.The recipient was given medical treatment, however, the issues did not resolve.Programming adjustments were made, however, the issue did not resolve.Device testing revealed results within normal limits.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: section d.9.Correction: section h.6.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The company was informed that the explanted device will not be returned to the company for analysis.A review of the device history record noted no anomalies.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey o'linn
28515 westinghouse place
valencia, CA 91355
MDR Report Key12825088
MDR Text Key280872014
Report Number3006556115-2021-01755
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016878046
UDI-Public(01)07630016878046(11)200928(17)230930
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/30/2023
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/16/2021
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
Supplement Dates FDA Received01/19/2022
01/31/2023
03/08/2023
05/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-