|
Model Number TN, FC500 FLOW CYTOMETER 100-240V, 50/60HZ |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/22/2021 |
Event Type
malfunction
|
Event Description
|
The customer reported amp board errors at the fl4 location of their fc 500 flow cytometer.There was no report of death, injury, or change to patient treatment as a result of this event.
|
|
Manufacturer Narrative
|
Section a patient information is not applicable.There was no impact to patients as a result of this event.The field service engineer (fse) was at the customer site and confirmed that the tarpon amp board at the fl4 channel location was functioning as expected the fse tested the instrument using flow-check and reported that channels fl1, fl2, fl3, fl4 and fl5 were stable but signal gaps persisted at the ss channel location.The fse replaced the tarpon amp board at the ss location.Bec is filing an mdr for this event based on the fda classification of the (b)(6) 2018 urgent medical device recall as a class i recall on (b)(6) 2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier case: (b)(4).
|
|
Search Alerts/Recalls
|
|
|