ZIMMER BIOMET SPINE INC. CURVED THORACIC PROBE Back to Search Results
Catalog Number 14-510151
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2021
Event Type  malfunction  
Event Description
It was reported that the tip of a pedicle probe broke during surgery.The piece was recovered without patient impacts.
 
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Manufacturer Narrative
Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation product was not returned and photos were not provided, so device evaluation could not be performed.Potential cause root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review per dhr review, the part was likely conforming when it left zimmer biomet control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event Description
It was reported that the tip of a pedicle probe broke during surgery.The piece was recovered without patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURVED THORACIC PROBE
Type of Device
PROBE
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key12828698
MDR Text Key280884712
Report Number3012447612-2021-00435
Device Sequence Number1
Product Code HXB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14-510151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/22/2021
Initial Date FDA Received11/17/2021
Supplement Dates Manufacturer Received03/18/2022
Supplement Dates FDA Received03/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-