Model Number 134031 |
Device Problems
Device Slipped (1584); Detachment of Device or Device Component (2907)
|
Patient Problem
Unspecified Tissue Injury (4559)
|
Event Date 11/02/2021 |
Event Type
Injury
|
Event Description
|
According to the reporter, during an open vein bypassing, when clipping the vessel, the surgeon was able to squeeze the handle, there were five unstable clips on the vessel that fell into the patient's cavity and were retrieved with an instrument.The vessels were covered by sutures to resolve the issue as standard procedure.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during an open vein bypassing, when clipping the vessel, the surgeon was able to squeeze the handle.There was one unstable clip from the first device on the vessel that fell into the patient's cavity and was retrieved with an instrument.It was noted that a second clip from a second product was also unstable and was tested on paper.The vessels were covered by sutures to resolve the issue as standard procedure.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during an open vein bypassing, when clipping the vessel, the surgeon was able to squeeze the handle, there was one unstable clip on the vessel that fell into the patient cavity and were retrieved with an instrument.It was noted that a second clip was also unstable and was tested on paper.The vessels were covered by sutures to resolve the issue as standard procedure.
|
|
Manufacturer Narrative
|
Medtronic conducted an investigation based upon all information received.Forty-two sealed representative samples with the appropriate packaging were available for evaluation.Visual inspection noted that the instrument was sealed with the full complement of clips.Functionally, the instrument was first test fired once in air so that the clip formation could be observed.The jaw and handle moved smoothly through the firing cycle and returned to the open position.It was reported that the clip did not hold tightly to the vessel and detached, a component disengaged from the device into the surgical cavity, and significant surgical intervention was required due to the product failure.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|