DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problems
Pumping Stopped (1503); Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Type of investigation not yet determined(4118) a supplemental report will be submitted when additional information is provided and/or our investigation is completed.The full name of the event site was shortened due to field character limit; the full name is (b)(6) medical center.Not returned to manufacturer.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) stopped and alarmed for low vacuum.Customer rebooted unit then it alarmed for an electrical code failure.Consequently, customer took pump to their biomed department to be evaluated.There was a slight delay in therapy as customer was switched to another cs300 iabp, however there was no harm to the patient and no adverse event reported.
|
|
Manufacturer Narrative
|
Additional information: e1 (event site telephone: (b)(6) a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the motor control drive and drive manifold, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
|
|
Event Description
|
N/a.
|
|