Model Number 0998-00-3023-53
Device Problems Pumping Stopped (1503); Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2021
Event Type  malfunction  
Manufacturer Narrative
Type of investigation not yet determined(4118) a supplemental report will be submitted when additional information is provided and/or our investigation is completed.The full name of the event site was shortened due to field character limit; the full name is (b)(6) medical center.Not returned to manufacturer.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) stopped and alarmed for low vacuum.Customer rebooted unit then it alarmed for an electrical code failure.Consequently, customer took pump to their biomed department to be evaluated.There was a slight delay in therapy as customer was switched to another cs300 iabp, however there was no harm to the patient and no adverse event reported.
 
Manufacturer Narrative
Additional information: e1 (event site telephone: (b)(6) a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the motor control drive and drive manifold, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12837883
MDR Text Key281926377
Report Number2249723-2021-02681
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAQUET IAB CATHETER
Patient Age60 YR
Patient SexMale
Patient Weight130 KG
-
-