Model Number 22330 |
Device Problem
Sensing Intermittently (1558)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
|
|
Event Description
|
The customer reported the cable receptacle is intermittent.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.During functionality testing, the unit powered on and off using battery power but a sensor was unable to connect to the device to take measurements.The mini d 20 connector of the instrument board was damaged.The device cannot take measurements due to bent sensor connector not allowing a proper connection.A service history record review reveals that this unit was in the field for over eight (8) years with no previous reported issues related to this reported event.
|
|
Event Description
|
The customer reported the cable receptacle is intermittent.No patient impact or consequences were reported.
|
|
Search Alerts/Recalls
|