Model Number 22330
Device Problem Sensing Intermittently (1558)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
 
Event Description
The customer reported the cable receptacle is intermittent.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned device was evaluated.During functionality testing, the unit powered on and off using battery power but a sensor was unable to connect to the device to take measurements.The mini d 20 connector of the instrument board was damaged.The device cannot take measurements due to bent sensor connector not allowing a proper connection.A service history record review reveals that this unit was in the field for over eight (8) years with no previous reported issues related to this reported event.
 
Event Description
The customer reported the cable receptacle is intermittent.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-5 PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key12840889
MDR Text Key281437296
Report Number3019388613-2021-00325
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K033296
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22330
Device Catalogue Number9196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/06/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-