• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES(IRELAND) MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES(IRELAND) MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the lock on the mayfield modified skull clamp (a1059) is broken and the ratchet has too much play side to side.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: failure analysis - unit received with the lock having rotational and lateral movement and a residue buildup was present.The movement of the ratchet extension was within tolerance.General maintenance performed and new components were added to replace worn internal parts.Root cause - complaint confirmed via inspection of the unit.The lock had rotational and lateral movement and a residue buildup.Unit required cleaning and replacement of worn parts due to routine wear and tear.Unit is beyond integra's 7 years recommended life cycle (manufactured in 2009).However, the movement of the ratchet extension was within tolerance.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
sragh, tullamore, co.offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12846124
MDR Text Key286062806
Report Number3004608878-2021-00638
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Initial Date Manufacturer Received 10/28/2021
Initial Date FDA Received11/19/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-