MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A Back to Search Results
Model Number A1059
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Mayfield modified skull clamp (a1059) was returned for evaluation.Device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: unit received with the lock having rotational and lateral movement and a residue buildup was present.The unit was received without the plunger cap.New components were added to replace worn internal parts.General maintenance and cleaning were performed.Missing parts and worn parts were replaced.Root cause analysis: complaint confirmed via inspection of the unit.The unit was received without the plunger cap.The lock had movement and required replacement of worn internal parts from routine wear and tear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that the threaded cap on mayfield modified skull clamp (a1059 ) was missing.It is unknown if there was patient involvement however; no patient injury or surgical delay has been reported.After the device was returned by the customer, evaluation showed that the lock has rotational and lateral movement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12846375
MDR Text Key283790757
Report Number3004608878-2021-00642
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2021
Date Manufacturer Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown