BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number MMC9638
Device Problems Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that five (5) continu-flo sets were ¿found to be protruding through the packaging once the box was opened, rendering them not sterile¿.This issue was identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual devices were not available; however, retention samples were evaluated.Visual inspection of the retention samples was performed with no issues noted.The retention samples were gravity tested then leak tested under water with no issues noted.The reported condition was not verified on the retention samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12848972
MDR Text Key282207654
Report Number1416980-2021-06883
Device Sequence Number1
Product Code FPA
UDI-Device Identifier05413760282671
UDI-Public(01)05413760282671
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMMC9638
Device Lot Number21B12T427
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/27/2021
Initial Date FDA Received11/19/2021
Supplement Dates Manufacturer Received12/16/2021
Supplement Dates FDA Received12/27/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-