Model Number UHI-4 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is ongoing; therefore, the root cause of the reported issue/malfunction cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
|
|
Event Description
|
The regional repair center was informed that a customer high flow insufflation unit was returned due to a report of "check gas insufflator, does not work" during a surgical intervention.The intended procedure was completed using the same device but by turning the unit on and off several times.There was no delay and no adverse outcome to the patient.The device was returned for service.Upon inspection and testing, the device emits a permanent audio beep after it has started and an e03 error occurs - excessive pressure when inflated due to a defective (cr) board and mainboard.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it was confirmed that alarm goes off when e03 is reported, and the display of the front panel disappears.No defect inside the device other than the phenomenon pointed out.It was surmised that the phenomenon occurred due to the defective pressure sensor mounted on the main board.The cause of the defective pressure sensor could not be identified.It was recommended to replace the main board.The product standards related to the phenomena was checked and found the following descriptions: classification: 12.Quality standards quality item: 7.Hardware failure quality characteristic: 1.Detecting hardware failure standard: 1.If the following phenomena occurred, hardware failure occurred in order to secure safety.(1) control circuit power voltage failure.(2) standard voltage failure of ad converter.(3) pressure sensor data failure.(4) pressure sensor initialization failure.(5) backup data sum check failure.The instruction manual identifies the following related verbiage: chapter 7.1 ¿troubleshooting guide.¿ ·problem: all leds are turned off.·possible cause: an error is detected in the internal circuitry of the high-flow insufflation unit.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|