EUROSETS S.R.L. ADVANCED MEMBRANE GAS EXCHANGE PMP STERILE; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number US5062
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/23/2021
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that while priming an extracorporeal membrane oxygenation (ecmo) circuit with a eurosets oxygenator, the oxygenator made a high pitch squealing noise while flowing at 2400 rpm.The higher the rpm, the higher the pitch and louder the noise.The oxygenator was removed from the circuit, and was replaced with another eurosets oxygenator.The noise resolved after the exchange.The oxygenator was never used on a patient.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: eurosets confirmed an audible flowing noise upon performing a pressure drop test to the returned eurosets amg (advanced membrane gas exchange) pmp (polymethylpentene) oxygenator.A specific root cause for the noise could not be conclusively determined through this evaluation.The eurosets oxygenator, lot # 7081108, was returned to abbott and an initial visual inspection was performed.Visual inspection of the oxygenator revealed that the oxygenator had segments of tubing connected to the blood inlet and outlet ports that were clamped and the oxygenator was filled with a priming solution.No obvious damage or other abnormalities were identified.The oxygenator was forwarded to the external manufacturer (eurosets) for technical analysis.The oxygenator was set up and a pressure drop test was performed with flow noising clearly audible.Eurosets reviewed the production documentation for the oxygenator lot and confirmed all tests from the production process were compliant with the technical specifications.Eurosets confirmed that they apply 100% production process tests to all of their devices.Eurosets is still investigating on the root cause of this kind of issue.The eurosets amg pmp instructions for use (ifu) is currently available.Under the list of warnings, the ifu warns that during the extracorporeal circulation (ecc) a backup oxygenator is necessary and also warns that the extracorporeal circulation has to be carefully and continuously checked.Also, under the list of warnings, the ifu warns that ¿before using the product it is advisable to carefully inspect it.Shipping and handling could cause structural and functional damage to the device.¿ the section entitled ¿oxygenator replacement¿, this document states that a spare oxygenator must always be available during perfusion.After 6 hours of use with blood or if particular situations occur, which may lead the person responsible for perfusion to determine the safety of the patient may be compromised (insufficient oxygenator performance, leaks, abnormal blood parameters, etc.), follow the procedure outlined in the ifu for oxygenator replacement.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED MEMBRANE GAS EXCHANGE PMP STERILE
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EUROSETS S.R.L.
strada statale 12, n°143
medolla, modena 41036
IT  41036
Manufacturer (Section G)
EUROSETS S.R.L.
strada statale 12, n°143
medolla, modena 41036
IT   41036
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key12851874
MDR Text Key287681323
Report Number3003752502-2021-00012
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier08034013782020
UDI-Public08034013782020
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141492
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model NumberUS5062
Device Catalogue NumberUS5062
Device Lot Number7081108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-