Model Number 176657 |
Device Problems
Mechanical Problem (1384); Device Slipped (1584)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure while clipping over vessel in a laparoscopic procedure, the device was not clipping correctly.To resolve the issue the surgeon used an alternative product.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument noted that the handle of the instrument was out of its home position.The condition was replicated by partial compression of the handle.It was reported that the clips did not close completely and were malformed.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when the handle is not squeezed completely as far as it will go.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to squeeze the handle as far as it will go may result in clip malformation and possible bleeding or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure while clipping over vessel in a laparoscopic procedure, the device was not clipping correctly.The clip was not clipping together properly and did not form correctly when trying to close the clip.To resolve the issue the surgeon used an alternative product.There was no patient injury.
|
|
Search Alerts/Recalls
|
|
|