Catalog Number A103200000000 |
Device Problems
Defective Device (2588); Detachment of Device or Device Component (2907); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/08/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that device will not hold file and noisy; check device.
|
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.Additional information regarding the patient outcome has been requested and will be submitted as it becomes available.
|
|
Manufacturer Narrative
|
Evaluation found the cartridge and head cap needed to be replaced due to wear.
|
|
Event Description
|
Additional information was received indicating that no injury resulted.
|
|