Model Number PN-004 075
Device Problems Fluid/Blood Leak (1250); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.A short circuit was detected between electrode 1 and thermocouple 2 (tc2).Dried saline was noted within the electrical connector, consistent with the short circuit detected.Further testing revealed a gap in the adhesive between the irrigation tube and the proximal tube inside the luer lock, allowing fluid to flow outside of the irrigation tubing and into the proximal tubing towards the electrical connectors, consistent with the dried saline noted within the electrical connector.
 
Event Description
This report is to advise of an event observed during analysis confirming an irrigation leak and short circuit of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12863904
MDR Text Key281205930
Report Number9680001-2021-00110
Device Sequence Number1
Product Code OAE
UDI-Device Identifier07640157990040
UDI-Public07640157990040
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/25/2021
Device Model NumberPN-004 075
Device Catalogue NumberPN-004075
Device Lot Number7300910
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-