Catalog Number VS-404
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Thrombosis/Thrombus (4440); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2021
Event Type  Injury  
Event Description
Physician was attempting to use the venaseal closure system during procedure to treat the great saphenous vein (gsv).Local anesthesia and transducer compression were used.Ifu was followed and a guidewire was used for insertion of the catheter.It was reported that the physician stated that venaseal got injected into the common femoral vein (cfv) and physician stated that there was a technique issue.The physician stated that there was not adequate compression at the junction.The product performed how it was supposed to perform.The vein is reported to have closed.The procedure was completed per ifu.The physician put the patient on blood thinners.The blood thinners were not part of the original treatment plan and were provided due to some of the venaseal getting into the cfv.Blood thinners were prescribed by doctor and were not over the counter medication.No patient injury reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key12867176
MDR Text Key281197348
Report Number9612164-2021-04535
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVS-404
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2021
Initial Date FDA Received11/24/2021
Supplement Dates Manufacturer Received02/03/2022
Supplement Dates FDA Received02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-